The selection of proper mounting position is the key for efficient running of measuring equipment in the long term. General requirements: good dustproof condition in door, environment clean and tidy; soft light and avoiding screen image dazzling; entrance control, changing shoes when coming in and out recommended. Details are specified below:
Temperature:18-23°C ,±1%/hr
Humidity:30-80% (non condensing)
vibration:less than 0.6g ,below 15HZ
Operating environment:15-30°C
Utility requirements:220V±5% ,50/60HZ ,1Φ ,___W , protective grounding wire,regulated power supply of UPS
Thickness of lab mat:100mm*100mm*(5mm-15mm) (in accordance with actual floor condition of customers)
Installation position:machine(left), operation desk(right)
Approximate space demand:500mm+ machine +100mm+ operation desk