竞技宝JJB·(中国)电子竞技平台 Operating Environment Instruction

    The selection of proper mounting position is the key for efficient running of measuring equipment in the long term. General requirements: good dustproof condition in door, environment clean and tidy; soft light and avoiding screen image dazzling; entrance control, changing shoes when coming in and out recommended. Details are specified below:


    Temperature:18-23°C ,±1%/hr


    Humidity:30-80% (non condensing)


    vibration:less than 0.6g ,below 15HZ


    Operating environment:15-30°C


    Utility requirements:220V±5% ,50/60HZ ,1Φ ,___W , protective grounding wire,regulated power supply of UPS


    Thickness of lab mat:100mm*100mm*(5mm-15mm) (in accordance with actual floor condition of customers)


    Installation position:machine(left), operation desk(right)


    Approximate space demand:500mm+ machine +100mm+ operation desk